OligoPool
PricingContactAbout
OligoPool

Free, professional tools for oligonucleotide design and analysis. Built for the molecular biology research community.

Contact OligoPool

Tools

  • Tm Calculator
  • GC Content Analyzer
  • Molecular Weight Calculator
  • Primer Analyzer
  • Batch Sequence QC
  • All 19 Tools

Guides

  • PCR Primer Design
  • Oligo Pool Synthesis
  • CRISPR Library Design
  • Melting Temperature
  • Oligo Quality Control
  • All Guides

Solutions

  • All Solutions
  • For Academic Labs
  • For Biotech
  • For Pharma R&D

Compare

  • vs IDT OligoAnalyzer
  • vs NEB Tm Calculator
  • Oligo Tool Suite
  • All Comparisons

Oligo Pools

  • Oligo Pool Design
  • How to Design
  • Vendor Decision

About

  • About Us
  • Contact
  • Pricing
  • Vendors
  • Resources
  • Privacy Policy
  • Terms of Service

© 2026 OligoPool.com. All rights reserved.

Partners:Oligo Pools
HomeGuidesDesign Illumina Adapters

Illumina Adapter Design for Dual Indexes and NGS QC

Use this page when you need to design or review Illumina-compatible adapters, choose between TruSeq and Nextera-style structures, decide when to use unique dual indexes, and run Tm plus structure checks before ordering. If you already know the next task, open Validate NGS Library Prep Oligos, Tm Calculator, GC Content Analyzer, Secondary Structure Predictor, Batch Sequence QC, and What QC Should I Request for Oligos?.

Key Takeaways

  • •Illumina-compatible molecules combine flow-cell binding regions, sequencing-primer sites, insert sequence, and index reads; the exact construct depends on the kit and run configuration.
  • •TruSeq-style ligation and Nextera/Illumina DNA Prep tagmentation use different attachment and primer assumptions. Follow the selected kit protocol rather than mixing structures by name.
  • •Index hopping and demultiplexing risk depend on platform chemistry, library cleanup, index design, and the run configuration. Use the current platform guidance to choose a mitigation.
  • •Unique or combinatorial dual indexing is a design choice tied to the platform, sample count, index set, and demultiplexing software; this page does not claim a universal risk percentage.
  • •Tm and structure outputs are screening evidence under entered assumptions. Confirm adapter compatibility, ligation conditions, and acceptance rules in the current library protocol.
  • •Before ordering, preserve required fixed regions, document index orientation, run sequence-level checks, and obtain vendor or core-facility confirmation for custom constructs.

What are you trying to decide?

→ Which adapter regions must stay fixed for Illumina compatibility?→ Should I use TruSeq-style ligation or Nextera-style adapters?→ Which dual-indexing strategy fits my run?→ How do I review index-hopping risk on NovaSeq or NextSeq?→ Which Tm or dimer checks should I run before ordering?→ Can I mix legacy and new Illumina UD index kits?

Related adapter checks

→ Open the NGS library prep validation guide→ Check adapter duplex Tm in your buffer→ Check adapter and index GC balance→ Screen adapters for hairpins and self-dimers→ Verify index sets and barcode spacing in batch→ Choose purification and vendor QC requirements→ Resolve Tm disagreements between calculators
Page contents
Which Parts of an Illumina Adapter Must Stay Fixed?Should You Use TruSeq or Nextera-Style Adapters?Which Dual Indexing Strategy Fits Your Run?How Do You Review Index-Hopping Risk on Illumina Platforms?Which QC Checks Matter Before Ordering Custom Adapters?What Changed in Illumina Index Kits in 2025-2026?ReferencesFrequently Asked Questions

Which Parts of an Illumina Adapter Must Stay Fixed?

One common Illumina-compatible layout is shown below (5' → 3' of the top strand). Treat it as a conceptual map and reconcile every region with the selected kit and run configuration:

// Conceptual adapted-library map; verify the kit-specific layout
5'- [flow-cell region] — [read primer site] — [i5] — [insert] — [i7] — [read primer site] — [flow-cell region] -3'
// Functional regions:
Flow-cell regions: preserve the sequences required by the selected platform document
Read primer sites: match the library kit and read orientation
i5 / i7: use the documented index orientation and read lengths

Why must P5 and P7 stay unchanged?

Flow-cell binding regions are complementary to oligos attached to the flow cell surface. During cluster generation, denatured library molecules hybridize to these surface oligos, initiating the platform-specific cluster-generation process. Their exact sequence, orientation, and adjacent primer sites are kit-dependent; preserve the regions required by the current Illumina document rather than copying a generic layout into every custom design.

Sequence5' → 3'Length in this exampleGC%Model output under stated assumptions
P5AATGATACGGCGACCACCGAGA22 nt54.5%~67°C
P7CAAGCAGAAGACGGCATACGAGAT24 nt50.0%~68°C

Any example sequence lengths and Tm values are calculation context, not universal kit specifications. Recalculate the complete duplex with the actual concentrations, ions, and method using our Tm Calculator.

Should You Use TruSeq or Nextera-Style Adapters?

FeatureTruSeqNextera / Illumina DNA Prep
FragmentationMechanical (sonication) or enzymaticTagmentation (Tn5 transposon)
Adapter attachmentLigation to dA-tailed fragmentsTransposon inserts partial adapters; indexes added by PCR
Adapter structureY-shaped (forked) — pre-formed P5/P7 duplex with single-stranded overhangMosaic End (ME) sequences — minimal transposon recognition + PCR-added indexes
Input DNAUse the selected kit protocolUse the selected kit protocol
Hands-on timeProtocol and workflow dependentProtocol and workflow dependent
Insert size rangeUser-defined (based on fragmentation)Protocol and size-selection dependent
Sequencing primer sitesTruSeq Read 1/2 primersNextera Read 1/2 primers (different sequences)

Compatibility note: do not mix TruSeq and Nextera libraries on the same lane

TruSeq and Nextera-style libraries can use different sequencing-primer assumptions. Before pooling, verify primer compatibility, index orientation, sample-sheet settings, and any custom-primer loading requirements with the selected instrument and sequencing core.

Which Dual Indexing Strategy Fits Your Run?

Illumina dual indexing uses two barcode reads (commonly described as i7/Index 1 and i5/Index 2) according to the platform's orientation and sample-sheet rules. The choice between combinatorial and unique pairs affects how a run handles index reuse and possible hopping; use the current platform guidance.

When is combinatorial dual indexing still acceptable?

In combinatorial dual indexing, individual i5 and i7 sequences can be reused while the pair identifies a sample. Whether that is acceptable depends on the platform, index set, sample sheet, and contamination tolerance; do not infer suitability from sample count alone.

When should you use unique dual indexing?

In unique dual indexing, each planned sample receives a non-reused i5+i7 pair within the pool. This can make some hopped combinations easier to identify, but the outcome depends on the index set, platform, sample sheet, controls, and demultiplexing software. Verify the current run documentation.

StrategyDecision variableConfirm before useCapacity
Single indexingIndex reuse and run riskPlatform and demultiplexing rulesIndex-set dependent
Combinatorial dual (CDI)Reusable component sequencesCurrent platform note and sample sheetIndex-set dependent
Unique dual (UDI)Non-reused pair within the poolCompatibility and demultiplexing rulesIndex-set dependent

How Do You Review Index-Hopping Risk on Illumina Platforms?

Index hopping (also called index switching or barcode swapping) occurs when index sequences from one library molecule become associated with a different library molecule, causing mis-assignment of sequencing reads to the wrong sample. This is a significant concern on patterned flow cell platforms.

Why does hopping happen on patterned flow cells?

Some patterned-flow-cell workflows use ExAmp chemistry, where free adapter or incomplete-library molecules can contribute to index misassignment. The observed effect depends on platform, cleanup, library mix, controls, and analysis settings; use the current platform technical note for the run.

Which mitigations matter most?

StrategyPurposeImplementation
Use UDI indexesReduce ambiguity from index reuseConfirm index-set and demultiplexing support
Clean up free adaptersRemove free adapters and dimersFollow the selected kit cleanup ratio
Minimize adapters in poolKeep library preparation within protocol scopeUse the current quantification and loading instructions
Fresh poolingControl storage and handling variablesUse the current kit and core-facility instructions

Which QC Checks Matter Before Ordering Custom Adapters?

Whether using standard adapters or designing custom ones, sequence and structure checks are screening steps. They do not establish ligation efficiency, flow-cell compatibility, or dimer performance without the selected protocol and a measured validation plan.

Which Tm ranges should adapter regions hit?

Adapter RegionHow to set the Tm checkRationale
Y-adapter duplex regionUse the selected duplex and buffer assumptionsCompare with the kit protocol, not a universal range
Flow cell binding (P5/P7)Preserve the platform document's required sequenceDo not infer cluster behavior from Tm alone
Sequencing primer bindingMatch the current sequencing-primer protocolConfirm orientation and chemistry with the platform
Index sequences (8 bp)Use the index-set specificationCheck read orientation, color balance, and demultiplexing rules

Which structures should force a redesign?

Custom adapter oligos must be evaluated for self-complementarity and hairpin formation. Problematic secondary structures reduce ligation efficiency and can cause adapter-dimers that consume sequencing capacity.

Structure TypeInterpretation ruleRisk LevelAction
HairpinUse the selected structure-policy bandScreening resultReview in assay context
HairpinCompare with the selected method and temperatureReview neededDo not infer ligation failure from ΔG alone
Self-dimerUse the selected structure-policy bandScreening resultReview in assay context
Self-dimerConfirm the selected structure-policy bandHigher review priorityRedesign or validate only with protocol evidence

Validate adapter oligos using our Secondary Structure Predictor for hairpin and self-dimer analysis, and our Tm Calculator to verify duplex stability at your ligation buffer conditions. For batch validation of custom adapter pools, use Batch Sequence QC.

What Changed in Illumina Index Kits in 2025-2026?

Product names and index sets change over time. The following notes are a dated source lead, not a guarantee that the current catalog or transition dates are unchanged:

  • Product naming: verify the current product name and document version before ordering.
  • Index revisions: check whether the selected index set has a revised sequence, orientation, or color-balance requirement.
  • Transition dates: treat any legacy-order deadline as source-specific and verify it directly with Illumina before planning a run.
  • Mixing sets: confirm compatibility between index-set versions, sample sheets, and the demultiplexing software before combining them.

The current Illumina adapter/index document and sequencing-core instructions are the authority for sequence, orientation, color balance, and version compatibility. Verify them immediately before library preparation.

References

  • Illumina. (2025). Illumina Adapter Sequences Document. Document #1000000002694.
  • Illumina. (2024). Index Hopping Filter — Understanding and Mitigating Index Cross-Talk. Illumina Technical Note.
  • Costello M, et al. (2018). Characterization and remediation of sample index swaps by non-redundant dual indexing on massively parallel sequencing platforms. BMC Genomics, 19:332.doi:10.1186/s12864-018-4703-0
  • SantaLucia J Jr. (1998). A unified view of polymer, dumbbell, and oligonucleotide DNA nearest-neighbor thermodynamics. PNAS, 95(4):1460-1465.
  • Adey A, et al. (2010). Rapid, low-input, low-bias construction of shotgun fragment libraries by high-density in vitro transposition. Genome Biology, 11:R119.doi:10.1186/gb-2010-11-12-r119

Frequently Asked Questions

What are P5 and P7 adapter sequences?

P5 and P7 are flow-cell binding regions used by Illumina-compatible library molecules. The exact sequences, orientation, and adjacent primer/index regions must be checked against the current kit or platform document; do not assume one layout applies to every library-prep product.

What is the difference between combinatorial and unique dual indexing?

Combinatorial dual indexing reuses individual i5 and i7 sequences across samples, while unique dual indexing assigns a non-reused pair within the planned pool. The practical benefit depends on platform chemistry, index design, sample sheet, and demultiplexing rules; verify the current instrument guidance before selecting one.

Why does index hopping occur on NovaSeq but not MiSeq?

Index hopping is associated with library and platform conditions that allow an index sequence to be transferred or misassigned. Patterned-flow-cell ExAmp workflows are a known case, but the observed rate depends on the run, cleanup, library mix, and analysis. Use the current platform technical note and run controls instead of a universal percentage.

How do I design custom Illumina-compatible adapters?

Preserve the fixed flow-cell and sequencing-primer regions required by the selected kit, place i5/i7 in the documented orientation, and validate the complete construct under the chosen method. Use the Tm Calculator and Secondary Structure Predictor as screening tools, then confirm the custom design with the current platform or core-facility protocol.

What causes adapter-dimer contamination and how should I reduce the risk?

Adapter dimers are products formed without the intended insert and can consume sequencing capacity. Their size, abundance, and cleanup depend on the kit and library. Follow the current adapter:insert, cleanup, input, and fragment-analysis instructions for the selected protocol; do not transfer a fixed ratio or size from another kit.

Which indexing strategy should I use for my experiment?

Choose the index strategy from the platform, library kit, sample count, index compatibility, and demultiplexing plan. Confirm whether the selected index sets can be mixed, how i5 is read, and which sample-sheet version is required. A current product document or sequencing core instruction is the authority for the run.

Related NGS Adapter Pages

Validate NGS Library Prep Oligos

Use this guide to check adapters, indexes, primers, and probes before ordering.

What QC Should I Request for Oligos?

Choose purification, read vendor QC, and set acceptance checks for custom adapters or probes.

Why Do Tm Calculators Disagree?

Match Tm methods and buffer assumptions before validating adapter duplexes or primer tails.