Trying to understand why the same primer gets a different Tm after you change Na⁺, Mg²⁺, or dNTP assumptions? This page isolates the salt-correction layer so you can compare how the formulas respond to the selected ion assumptions and see why two calculators can disagree even when they use the same nearest-neighbor thermodynamics.
All Tm values are calculated using the SantaLucia (1998) nearest-neighbor thermodynamic parameters as the base method. The four salt correction approaches are applied independently to the same base Tm:
Conditions: 250 nM oligo concentration, 0% DMSO, non-self-complementary sequences. Treat these as the comparison component's fixed assumptions. Calculations run in your browser by default. Dashboard History can store the full input snapshot when auto-save, session, or account history is enabled; Library saves and membership/API workflows can send sequence assets by design.
Sequence: ATCGATCGATCGATCGATCG (20 nt)
| Na⁺ (mM) | Owczarzy 2008 | SL98 Salt | von Ahsen 2001 | %GC Method | Max Spread |
|---|---|---|---|---|---|
| 10 | 38.45°C | 46.59°C | 38.45°C | 38.8°C | ±8.1°C |
| 25 | 47.23°C | 50.72°C | 47.23°C | 45.4°C | ±5.3°C |
| 50 | 53.09°C | 53.93°C | 53.09°C | 50.4°C | ±3.5°C |
| 75 | 56.16°C | 55.83°C | 56.16°C | 53.3°C | ±2.9°C |
| 100 | 58.17°C | 57.19°C | 58.17°C | 55.4°C | ±2.8°C |
| 150 | 60.75°C | 59.13°C | 60.75°C | 58.3°C | ±2.5°C |
| 200 | 62.4°C | 60.52°C | 62.4°C | 60.4°C | ±2.0°C |
| 300 | 64.44°C | 62.5°C | 64.44°C | 63.3°C | ±1.9°C |
Method observation: At standard PCR conditions (50 mM Na⁺), Owczarzy 2008 and SantaLucia 1998 Salt methods usually stay close for these example sequences. The %GC method can diverge by several degrees, so it should be used only for rough estimates. The largest disagreement between NN methods occurs at extreme salt concentrations (>200 mM Na⁺ or >4 mM Mg²⁺), where Owczarzy 2008 accounts for competitive ion binding that older formulas ignore.
| Method | Best For | Limitation | Used By |
|---|---|---|---|
| Owczarzy 2008 | PCR with Mg²⁺ + Na⁺/K⁺ mixed buffers | Requires both ion concentrations | NEB, OligoPool |
| SantaLucia 1998 Salt | Na⁺-only buffers (hybridization, FISH) | Does not account for Mg²⁺ effect | Primer3, early IDT |
| von Ahsen 2001 | Clinical diagnostics (standardized buffers) | Less validated for extreme GC% | Roche LightCycler |
| %GC Method | Quick mental estimates only | Ignores sequence context, ±5-10°C error | No modern tool |